Article type: Research Article
Authors: van der Eijk, A.C.; | van der Plas, A.J. | van der Palen, C.J.N.M. | Dankelman, J. | Smit, B.J.
Affiliations: Department of BioMechanical Engineering, Faculty of Mechanical, Maritime & Materials Engineering, Delft University of Technology, Delft, The Netherlands | Department of Medical Technology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands | Department of Neonatology, Erasmus MC, Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, The Netherlands | Patient Directorate Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands
Note: [] Corresponding author: Dr. Anne Catherine van der Eijk, Department of Neonatology, Erasmus MC, Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, Dr. Molewaterplein 60, 3015 GJ Rotterdam, The Netherlands. Tel.: +31 010 7037008; Fax: +31 010 7036811; E-mail: [email protected]
Abstract: BACKGROUND: In multi-infusion IV therapy, the actual volume delivered to the neonate can vary over time. To reduce flow rate variability, check valves can be used. A check valve allows flow through the valve in only one direction. OBJECTIVE: To evaluate flow rate variability in a low flow dual-infusion setup with and without check valves. METHODS: The effect of changing the height of and adding syringes to the IV-administration set was tested with and without check valves in an in vitro dual-infusion setup with in-line flow meters. The pre-programmed flow rates were 2.5 and 0.1 ml/h. RESULTS: Twenty-four tests of 90 minutes were performed. Time to reach 75% of the pre-programmed 0.1 ml/h flow rate was >20 minutes. The highest total delivered volume during a test was (mean ± SD) 56 ± 8% of the expected delivery for tests without check valves, and diminished to 12 ± 24% of the expected delivery for check valves with a higher opening pressure. CONCLUSIONS: The actual flows and the total delivered volume in low flow dual-infusion setups are less than expected on the pre-programmed flow-rate. These findings emphasize the need for the development of more accurate delivery systems for drugs and fluids in neonatology. Caregivers should be aware of these findings, and optimise the delivery of IV substances by making use of check valves with low opening pressures and by minimising compliance and volume of the IV-administration set. Furthermore, changes in the relative height between pumps and catheter tip should be minimized.
Keywords: Neonatology, multi-infusion, IV therapy, check valve, anti-siphon
DOI: 10.3233/NPM-1475213
Journal: Journal of Neonatal-Perinatal Medicine, vol. 7, no. 1, pp. 55-64, 2014
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]