Article type: Research Article
Authors: El-Kafrawy, Ulaa; * | Taylor, R.J.b
Affiliations: [a] Neonatal Intensive Care Unit, Royal Bolton Hospital, Bolton, UK | [b] Department of Medical Physics, Salford Royal NHS Foundation Trust, Salford, UK
Correspondence: [*] Address for correspondence: Ula El-Kafrawy, MD, Department of Neonatology, Royal Bolton Hospital, Minerva Road, Bolton, BL4 0JR, UK. Tel.: +44 204 390335/07815197407; Fax: +44 204 390211; E-mail: [email protected].
Abstract: OBJECTIVES: To determine the accuracy of weighing ventilated infants on incubator scales and whether the accuracy can be improved by the addition of a ventilator tube compensator (VTC) device to counterbalance the force exerted by the ventilator tubing. STUDY DESIGN: Body weights on integral incubator scales were compared in ventilated infants (with and without a VTC), with body weights on standalone electronic scales (true weight). Individual and series of trend weights were obtained on the infants. The method of Bland and Altman was used to assess the introduced bias. RESULTS: The study included 60 ventilated infants; 66% of them weighed <1000 g. A total of 102 paired-weight datasets for 30 infants undergoing conventional ventilation and 30 undergoing high frequency oscillator ventilation (HFOV) supported by a SensorMedics oscillator, (with and without a VTC) were obtained. The mean differences and (95% CI for the bias) between the integral and true scale weighing methods was 60.8 g (49.1 g to 72.5 g) without and –2.8 g (–8.9 g to 3.3 g) with a VTC in HFOV infants; 41.0 g (32.1 g to 50.0 g) without and –5.1 g (–9.3 g to –0.8 g) with a VTC for conventionally ventilated infants. Differences of greater than 2% were considered clinically relevant and occurred in 93.8% without and 20.8% with a VTC in HFOV infants and 81.5% without and 27.8% with VTC in conventionally ventilated infants. CONCLUSIONS: The use of the VTC device represents a substantial improvement on the current practice for weighing ventilated infants, particularly in the extreme preterm infants where an over- or underestimated weight can have important clinical implications for treatment. A large-scale clinical trial to validate these findings is needed.
Keywords: Neonatology, intensive care, ventilated infants, incubator scales, weight measurement
DOI: 10.3233/NPM-161623
Journal: Journal of Neonatal-Perinatal Medicine, vol. 9, no. 4, pp. 377-383, 2016
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]