Affiliations: [a] Department of Molecular Neuroscience, Institute of Neurology, University College London, London, UK | [b] Department of Neurodegeneration, Dementia Research Centre, Institute of Neurology, London, UK | [c] Department of Neuroimmunology, Institute of Neurology, University College London, London, UK | [d] Department of Psychiatry and Neurochemistry, Institute of Neuroscience and Physiology, The Sahlgrenska Academy at the University of Gothenburg, Gothenburg, Sweden
Correspondence: [*] Correspondence to: Jamie Toombs, Department of Molecular Neuroscience, Institute of Neurology, Queen Square House, University College London, London, WC1N 3BG, UK. Tel.: +44 02034484204; E-mail: [email protected].
Abstract: The effect of spinal manometers on cerebrospinal fluid (CSF) amyloid-β (Aβ) concentration was investigated. Pooled human CSF samples were divided in two, one half passed through a manometer into a collection tube, the other transferred directly to a collection tube. CSF was analyzed for Aβ38/40/42 using an electrochemiluminescence immunoassay. Relative to control, use of a manometer decreased Aβ38/40/42 concentration by 5.6% (±1.5SE), 4.4% (±1.7SE), and 4.3% (±2.4SE), respectively. The ratios of Aβ42 :40, Aβ42 :38, and Aβ40 :38 were not affected by manometer treatment. Factors which artificially lower CSF Aβ concentrations are relevant to clinical diagnosis for AD and study design.
Keywords: Alzheimer’s disease, amyloid-β, biomarkers, cerebrospinal fluid, manometer, pre-analytical factor
DOI: 10.3233/JAD-161126
Journal: Journal of Alzheimer's Disease, vol. 56, no. 3, pp. 885-891, 2017
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]