Article type: Research Article
Authors: Materko, Wollnera; b; * | dos Reis Façanha, Caio Cesara | Guedes, Gizelly Coelhoa | Dias, Marcela Fabiani Silvaa | Costa, Alisson Vieirab | Belfort, Dilson Rodriguesb | da Pureza, Demilto Yamaguchia; b | Alberto, Alvaro Adolfo Duartea
Affiliations: [a] Master Program in Health Sciences, Universidade Federal do Amapá, Macapá, AP, Brazil | [b] Undergraduate Course in Physical Education, Universidade Federal do Amapá, Macapá, AP, Brazil
Correspondence: [*] Corresponding author: Wollner Materko, Universidade Federal do Amapá, Rodovia Osmar Chaves Pinto Km 02-Jardim Marco Zero Macapá, AP 68903197, Brazil. E-mail: [email protected].
Abstract: BACKGROUND: A Polar heart rate monitor is a device that measures RR intervals, but has not been correlated to accurately measure the series of RR interval signals between the ECG and the Polar® Heart Rate Monitor Interface (HRMI) Board at rest. OBJECTIVE: The aim of the study was to evaluate the temporal cross-correlations between the Polar® HRMI Board and an ECG to measure the series of RR intervals at rest. METHODS: The sample consisted of eighteen healthy male subjects and they were instructed to lie in the supine position at rest while breathing normally and a time window of the last 2 min was recorded to analyse RR intervals were obtained for each subject with a Polar® HRMI Board and an ECG. Cross-correlation analysis of RR interval signals between methods and reliability was expressed by Bland and Altman analysis. RESULTS: The cross-correlation was excellent, resulting in a mean of 0.98 ± 0.01 and no lag or delay between the signals. The bias was 0.03 ± 0.08 s or 8.0% for MeanRRi from Polar® HRMI Board and ECG, no significant difference. CONCLUSIONS: The Polar® HRMI Board is acceptable for assessment of serial RR intervals. The results support the reliability of the Mean RR interval compared to a resting ECG.
Keywords: Bland and Altman, heart rate variability, RR interval, electrocardiograph.
DOI: 10.3233/IES-230061
Journal: Isokinetics and Exercise Science, vol. 32, no. 1, pp. 59-64, 2024
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]