Authors: Dauty, Marca | Delbrouck, Clairea | Huguet, Dominiquec | Rousseau, Bertranda | Potiron-Josse, Michèleb | Dubois, Charlesa
Affiliations: [a] Pôle de Médecine Physique et Réadaptation, CHU Nantes, 44035 Nantes cedex 01, France | [b] Institut Régional de Médecine du Sport, CHU Nantes, 44035 Nantes cedex 01, France | [c] Polyclinique de l'Atlantique, Avenue Claude Bernard, 44 819 Saint Herblain, France
Note: [] Address for correspondence: Pôle de Médecine Physique et Réadaptation, Hôpital Saint Jacques, CHU Nantes, 44035 Nantes cedex 01, France. Tel.: +33 2 40 84 62 11; Fax: +33 2 40 84 61 91; E-mail: [email protected]
Abstract: The purpose of this study was to define the test-retest reproducibility of concentric and eccentric isokinetic measurements of shoulder rotators. It was carried out at a 26 day ± 4 interval. Fourteen healthy volunteer subjects (7 men and 7 women), aged between 40 and 55 years (47.5 ± 5.6 years), were included. Evaluation was carried out with a Cybex Norm dynamometer and the reliability was calculated according to the intra-class correlation coefficient (ICC 2.1). The protocol was randomised for the shoulder and for sets of concentric (con) and eccentric (ecc) tests. Concentric sets consisted of 5 movements of internal (IR) and external rotation (ER) at the speed of 60°/sec. then 120°/sec. Eccentric performance was recorded only for the external rotators, to determine the ratio ER_{ecc}/ IR_{con}. Results indicated excellent reproducibility for the peak torque (0.91–0.98), total work (0.88–0.96) and average power (0.83–0.94) in concentric muscle action for internal and external rotators and in eccentric muscle action for external rotators at 60 and 120°/sec. On the other hand the strength ratio: ER_{con} / IR_{con}and ER_{ecc} / IR_{con} should be interpreted with caution because of a lower reproducibility (ICC: 0.53–0.83). In conclusion, isokinetic findings relating to concentric and eccentric performance of the internal and external rotators of the shoulder are reproducible under the present protocol. These findings may serve for assessing rotator cuff function at the age where the incidence of rotator cuff overuse becomes a serious problem.
Keywords: shoulder, isokinetic, concentric, eccentric, reproducibility
DOI: 10.3233/IES-2003-0134
Journal: Isokinetics and Exercise Science, vol. 11, no. 2, pp. 95-100, 2003
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]