Article type: Research Article
Authors: Yu, Kaihonga | Yamashita, Tetsuib | Shingyochi, Shigeakic | Matsumoto, Kazuod | Komiya, Atsukia | Ohta, Makotoa; *
Affiliations: [a] Institute of Fluid Science, Tohoku University, Sendai, Miyagi, Japan | [b] Cardiovascular Internal Medicine System Sec., JMS Co., Ltd., Tokyo, Japan | [c] R&D Center, Nidec Copal Electronics Corp., Sano, Tochigi, Japan | [d] Cardiology Department, Saitama Medical University, Hidaka, Saitama, Japan
Correspondence: [*] Corresponding author: Makoto Ohta, Institute of Fluid Science, Tohoku University, 2-1-1 Katahira Aoba-Ku 980-8577, Sendai, Miyagi, Japan. Tel./Fax: +81 22 217 53 09; E-mail: [email protected].
Abstract: BACKGROUND: A new cooling method is proposed for preventing electrode-tissue overheating during cardiac catheter ablation using a vibrating catheter. Previous work has shown that vibration that results from increased flow velocity around the catheter can have a cooling effect on the electrode. OBJECTIVE: Contact force has been shown to be an important factor that affects cooling and lesion formation, because contact force determines the ratio of power delivery between blood and tissue. In this study, the effect of contact force on electrode cooling and tissue heating was investigated during the operation of an electrode cooled by vibration. METHODS: Using PVA-H or myocardium ablation tissue models under conditions of no flow, electrode and tissue temperatures and lesion sizes were measured at various vibrational frequencies and contact force conditions. RESULTS: The experiments showed that the catheter vibration still decreases the electrode temperature over a contact force range of 2–30 gf. The lesion size was increased with increasing contact force at each vibrational frequency. CONCLUSIONS: Increasing contact force can increase lesion size with cooling by vibration remaining effective.
Keywords: Catheter ablation, electrode cooling, contact force, vibration, PVA-H
DOI: 10.3233/THC-191639
Journal: Technology and Health Care, vol. 27, no. 6, pp. 589-601, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]