Abstract: Drug-eluting stent (DES) is tiny wire mesh tubes coated with drug used to open narrowed and diseased coronary arteries. DES releases drug slowly to prevent scar tissue growth in the artery, which causes in-stent restenosis. However, clinical problems remain such as stent fractures and polymer coating detachment. During the balloon catheter expansion, DES has to undergo the plastic deformation. Moreover, after implantation, stent has to undergo the load given by the heart pulse. In this study, fatigue behavior of stent is investigated and morphologies of polymer coating in relation to stent expansion and cyclic strain were observed using scanning electron microscope (SEM) to assess the reliability of DES. Fatigue test of stent indicated that there were no significant differences in fatigue behavior between DES and bare-metal stent (BMS). From the SEM appearance, two kinds of irregularities caused by expansion process were observed in the coating of DES, that is, crack and delamination. Additionally, no significant increase of irregularities was found in the coating of DES after fatigue test.
Keywords: , Polymer coated drug eluting stent, fatigue test, adhesion test
DOI: 10.3233/SFC-2011-0138
Journal: Strength, Fracture and Complexity, vol. 7, no. 2, pp. 195-203, 2011
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]