Article type: Research Article
Authors: Liu, Boa | Haithem Zaki, Shaimab; * | García, Eduardoc | Bonilla, Amandad | Thabit, Dahae | Hussein Adab, Ayaf
Affiliations: [a] Magnetic Resonance Imaging Room, Henan Province Hospital of Traditional Chinese Medicine, Zhengzhou, China | [b] Department of Anesthesia Techniques, Al-Noor University College, Nineveh, Iraq | [c] Facultad de Mecánica, Escuela Superior Politécnica de Chimborazo (ESPOCH), Riobamba, Ecuador | [d] Facultad de Ciencias, Escuela Superior Politécnica de Chimborazo (ESPOCH), Sede Orellana, El Coca, Ecuador | [e] Medical Technical College, Al-Farahidi University, Baghdad, Iraq | [f] Department of Pharmacy, Al-Zahrawi University College, Karbala, Iraq
Correspondence: [*] Corresponding author: Shaima Haithem Zaki, Department of Anesthesia Techniques, Al-Noor University College, Nineveh, Iraq. E-mails: [email protected].
Abstract: OBJECTIVE:It seems that dose rate (DR) and photon beam energy (PBE) may influence the sensitivity and response of polymer gel dosimeters. In the current project, the sensitivity and response dependence of optimized PASSAG gel dosimeter (OPGD) on DR and PBE were assessed. MATERIALS AND METHODS:We fabricated the OPGD and the gel samples were irradiated with various DRs and PBEs. Then, the sensitivity and response (R2) of OPGD were obtained by MRI at various doses and post-irradiation times. RESULTS:Our analysis showed that the sensitivity and response of OPGD are not affected by the evaluated DRs and PBEs. It was also found that the dose resolution values of OPGD ranged from 9 to 33 cGy and 12 to 34 cGy for the evaluated DRs and PBEs, respectively. Additionally, the data demonstrated that the sensitivity and response dependence of OPGD on DR and PBE do not vary over various times after the irradiation. CONCLUSIONS:The findings of this research project revealed that the sensitivity and response dependence of OPGD are independent of DR and PBE.
Keywords: Dose rate, optimized PASSAG, magnetic resonance imaging, photon beam energy, polymer gel dosimeter
DOI: 10.3233/XST-230282
Journal: Journal of X-Ray Science and Technology, vol. 32, no. 3, pp. 751-764, 2024
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]