Article type: Research Article
Authors: Meng, Wan-Binb | Li, Qianga; * | Ye, Yan-Chengc | Wu, Jia-Mingc; d; e; *
Affiliations: [a] Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou, China | [b] University of Chinese Academy of Sciences, Beijing, China | [c] Heavy Ion Center of Wuwei Cancer Hospital; Gansu Wuwei Academy of Medical Sciences, Gansu Wuwei Tumor Hospital, Wuwei City, Gansu Province, China | [d] Department of Medical Physics, Chengde Medical University, Chengde City, Hebei Province, China | [e] Department of Radiation Oncology, Yee Zen General Hospital, Tao Yuan City, Taiwan
Correspondence: [*] Correspondence author: Qiang Li, Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China. E-mail: [email protected] and Jia-Ming Wu, Heavy Ion Center of Wuwei Cancer Hospital, Gansu Wuwei Academy of Medical Sciences, Wuwei City, Gansu Province, China. Tel.: +82 188 7358 8300; Fax: +82 0735 2343382; [email protected].
Abstract: OBJECTIVE:This study aims to develop and test a new technique by using the convergent arcTAN (cATAN) method capable of dealing with the virtual source position delivered by different carbon ion energies from the pattern of scanning-passive scatter beam. MATERIALS AND METHODS:A homemade large-format CMOS sensor and Gaf Chromic EBT3 films are used for the virtual source position measurement. The Gaf films are embedded in a self-designed rectangular plastic frame to tighten films and set up on a treatment couch for irradiation in air with the film perpendicular to the carbon ion beam at the nominal source-axis-distance (SAD) as well as upstream and downstream from the SAD. The horizontal carbon ion beam with 5 energies at a machine opening field size is carried out in this study. The virtual source position is determined by using the convergent arcTAN (cATAN) method and compared with a linear regression by back projecting FWHM to zero at a distance upstream from the various source-film-distance. RESULTS:The film FWHM measurement error of 0.5 mm leads to 0.001% deviation of α (cATAN) at every assumed textend. The overall uncertainty for the reproducibility of calculated virtual source position by the assumed textend in the vertical and horizontal directions amounts to 0.1%. The errors of calculated virtual source position by assumed textend with back projecting FWHM to zero methods are within 1.1±0.001, p = 0.033. CONCLUSION:We develop a new technique capable of dealing with the virtual source position with a convergent arcTAN method to avoid any manual measurement mistakes in scanning-passive scatter carbon ion beam. The readers are encouraged to conduct the proposed cATAN method in this study to investigate the virtual source position in the Linac-based external electron beams and the proton beams.
Keywords: Carbon ion beams, virtual source position, scanning-passive scatter beam
DOI: 10.3233/XST-221274
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 1, pp. 181-197, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]