Article type: Research Article
Authors: Jeon, Pil-Hyuna | Lee, Chang-Laeb; *
Affiliations: [a] Department of Diagnostic Radiology, Yonsei University Wonju College of Medicine, Wonju Severance Christian Hospital, Wonju-si, Gangwon-do, Republic of Korea | [b] Health & Medical Equipment Business Unit, Samsung Electronics, Suwon-si, Gyeonggi-do, Republic of Korea
Correspondence: [*] Corresponding author: Chang-Lae Lee, Health & Medical Equipment Business Unit, Samsung Electronics 129, Samsungro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea. Tel.: +82 31 200 6890; Fax: +82 31 200 6723; E-mail: [email protected].
Abstract: BACKGROUND:Expanding computed tomography (CT) detector coverage broadens the beam width, but inaccurate tube current application can reduce image quality at the boundaries between body regions with different attenuation values along the z-axis. OBJECTIVE:This study aims to develop and validate a new CT scanning technique with a fixed pitch to achieve higher imaging quality. METHODS:A cylindrical water phantom and an anthropomorphic chest phantom with different diameters represent a human body with different attenuation values. By optimizing the beam width and helical pitch, the pitch is fixed during scanning. The mean noise of the images and the standard deviation were calculated, and the coefficient of variation (COV) was compared to evaluate the uniformity of image noise according to the beam width. RESULTS:At the boundaries between regions with different attenuation values, the 10 mm beam width (COV: 0.065) in the water phantom showed a 47.7% COV reduction of image noise compared with the 20 mm beam width (COV: 0.125). In addition, the 20 mm beam width (COV: 0.146) in the chest phantom showed a 29.3% COV reduction of image noise compared with the 40 mm beam width (COV: 0.206). Thus, as the beam was narrowed, the mean noise was similar, but the standard deviation was reduced. CONCLUSIONS:The proposed CT scanning technique with a fixed pitch, optimized beam width, and helical pitch demonstrates that image quality can be improved without increasing radiation dose at the boundary between regions with different attenuation values.
Keywords: X-ray beam width, computed tomography (CT) scanning, helical pitch, CT image noise, CT image quality.
DOI: 10.3233/XST-211103
Journal: Journal of X-Ray Science and Technology, vol. 30, no. 2, pp. 399-408, 2022
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]