Article type: Research Article
Authors: Wang, Zhongfeia; 1 | Yun, Qinghuib; 1 | Liu, Changhaoa | Sun, Xiaohuana | Wang, Weia | Yin, Yutiana | Xiao, Fenga; * | Zhao, Linaa; *
Affiliations: [a] Department of Radiation Oncology, Xijing Hospital, Air Force Medical University, Xi’an, Shaanxi Province, China | [b] Department of Equipment, Xijing Hospital, Air Force Medical University, Xi’an, Shaanxi Province, China
Correspondence: [*] Corresponding author: Lina Zhao, M.D., Ph.D., Department of Radiation Oncology, Xijing Hospital, Air Force Medical University, Xi’an, Shaanxi Province, China. Tel.: +86 13572553455; E-mail: [email protected] and Feng Xiao, B.S., Department of Radiation Oncology, Xijing Hospital, Air Force Medical University, Xi’an, Shaanxi Province, China. Tel.: +86 13991291156; E-mail: [email protected].
Note: [1] Zhongfei Wang and Qinghui Yun contributed equally to this work.
Abstract: OBJECTIVE:To improve safety and efficiency of radiotherapy process by customizing a Varian ARIA oncology information system following the guidelines provided in AAPM TG-100 report. METHODS:First, failure mode and effects analysis (FMEA) and quality management program were implemented for radiotherapy process. We have customized the visual care path in the ARIA system and set up a series of templates for simulation, prescription, contouring, treatment planning, and multiple checklists. Average time of activities’ completion and amount of planning errors were compared before and after the use of the customized ARIA to evaluate its impact on the efficiency and safety of radiotherapy. RESULTS:Completion time and on-time completion rate of the key activities in the care path are improved. The time of OAR/targets contouring decreases from (1.94±1.51) days to (1.64±1.07) days (p = 0.003), with the on-time completion rate increases from 77.4%to 83.3%(p = 0.048). Treatment planning time decreases from (0.81±0.65) days to (0.55±0.51) days (p < 0.001), with the on-time completion rate increases from 96.6%to 98.3%(p = 0.163). Waiting time of patients decreases from (4.50±1.83) days to (4.04±1.34) days (p < 0.001), with the on-time completion rate increases from 81.9%to 89.7%(p = 0.003). In addition, the average plan error rate decreases from 5.5%(2.9%for safety errors and 2.6%for non-normative errors) to 2.4%(1.6%for safety errors and 0.8%for non-normative errors) (p = 0.029). CONCLUSION:Our study demonstrates that the customized ARIA system has the potential to promote efficiency and safety in radiotherapy process management. It is beneficial to organize and accelerate the treatment process with more effective communications and fewer errors.
Keywords: TG-100, ARIA system, quality control, quality assurance
DOI: 10.3233/XST-210952
Journal: Journal of X-Ray Science and Technology, vol. 29, no. 6, pp. 1103-1112, 2021
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]