Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Harun, H.H.a | Karim, M.K.A.a; * | Abbas, Z.a | Sabarudin, A.b | Muniandy, S.C.c | Ibahim, M.J.d
Affiliations: [a] Department of Physics, Faculty of Science, Universiti Putra Malaysia, Serdang, Selangor, Malaysia | [b] Department of Diagnostic & Applied Health Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysiainstitution> | [c] Department of Radiology, Hospital Kuala Lumpur, Jalan Pahang, Kuala Lumpur, Malaysia | [d] Department of Biochemistry and Molecular Medicine, Faculty of Medicine, Universiti Teknologi MARA, Sungai Buloh, Selangor, Malaysia
Correspondence: [*] Corresponding author: M.K.A. Karim, Department of Physics, Faculty of Science, Universiti Putra Malaysia, 43400 Serdang, Selangor, Malaysia. E-mail: [email protected].
Abstract: PURPOSE:To evaluate the influence of iterative reconstruction (IR) levels on Computed Tomography (CT) image quality and to establish Figure of Merit (FOM) value for CT Pulmonary Angiography (CTPA) examinations. METHODS:Images of 31 adult patients who underwent CTPA examinations in our institution from March to April 2019 were retrospectively collected. Other data, such as scanning parameters, radiation dose and body habitus information from the subjects were also recorded. Six different levels of IR were applied to the volume data of the subjects. Five circles of the region of interest (ROI) were drawn in five different arteries namely, pulmonary trunk, right pulmonary artery, left pulmonary artery, ascending aorta and descending aorta. The mean Signal-to-noise ratio (SNR) was obtained, and the FOM was calculated in a fraction of the SNR2 divided by volume-weighted CT dose index (CTDIvol) and SNR2 divided by the size-specific dose estimates (SSDE). RESULTS:Overall, we observed that the mean value of CTDIvol and SSDE were 13.79±7.72 mGy and 17.25±8.92 mGy, respectively. Notably, SNR values significantly increase with increase of the IR level (p < 0.05). There are also significant differences (p < 0.05) in the FOM for both SNR2/SSDE and SNR2/CTDIvol attained in different IR levels. CONCLUSION:We successfully evaluate the value of radiation dose and image quality performance and set up a figure of merit for both parameters to further verify scanning protocols by radiology personnel.
Keywords: Signal-to-noise ratio, figure-of-merit, CT pulmonary angiography, pulmonary embolism, iterative reconstruction algorithm
DOI: 10.3233/XST-200699
Journal: Journal of X-Ray Science and Technology, vol. 28, no. 5, pp. 893-903, 2020
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]