Article type: Research Article
Authors: Kong, Que | Lu, Fengqi; * | Gao, Yu | Qiao, Peng | Shao, Min | Xie, Min
Affiliations: Department of Radiology, Nanjing Medical University Affiliated Wuxi Second Hospital, Wuxi, China
Correspondence: [*] Corresponding author: Fengqi Lu, Department of Radiology, Wuxi, Jiangsu Province 214002, China. E-mail: [email protected].
Abstract: BACKGROUND:Computed tomography (CT) is considered a standard modality for imaging the paranasal sinus (PS), but increasingly radiation dose is of concern, especially in children. Objective:This study aims to investigate the feasibility of using a 320-detector CT scanner with a 16 cm wide-detector combined with iterative reconstruction (IR) algorithm to further reduce radiation dose when scanning the PS. METHODS:A total of 90 children who underwent CT of the PS were randomly allocated into three groups namely, (1) the experimental group using low-dose wide-detector scan (n = 30, 9±4 years); (2) low-dose helical group (n = 30, 9±4 years); and (3) pediatric conventional group (n = 30, 8±4 years). Statistical software SPSS 19.0 was used for one-way ANOVA analysis of the general data (age, BMI), image quality, and radiation dose. Multiple comparisons of data without homogeneity of variance were analyzed by Bonferroni test and Tamhane’s test. RESULTS:All patients underwent successful CT examinations. No significant differences in the general data and image quality evaluation were detected between three groups (all P values > 0.05). CTDIvol and DLP were 2.87 mGy and 32.58 mGy·cm in the experimental group, 4.92 mGy and 70.84 mGy·cm in the low-dose helical group, and 9.95 mGy and 131.83 mGy·cm in the conventional group, respectively, which were significantly different among these three groups as indicated by multiple comparisons (all P values < 0.05). In the experimental group, the effective radiation dose was 0.07 mSv, which was reduced by 76% and 56% comparing to the conventional group and the low-dose helical group, respectively. CONCLUSIONS:The 320-detector CT scanner equipped with the wide-detector combined with IR can further reduce radiation dose, while maintaining good image quality comparing to the low-dose helical or pediatric modes.
Keywords: 320-detector computed tomography (CT), wide-detector, iterative reconstruction, pediatric, paranasal sinus, low dose
DOI: 10.3233/XST-180487
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 4, pp. 731-738, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]