Article type: Research Article
Authors: Brown, Lee E.a; *; ** | Whitehurst, Michaela; *** | Bryant, Jennifer R.a; ****
Affiliations: [a] American Med-Care Center and Rehabilitation Facility
Note: [*] Exercise Physiologist, South Palm Orthopedics, Delray Beach, FL.
Note: [**] Address reprint requests to Lee E. Brown at South Palm Orthopedics, 16244 S. Military Trail, Suite 590, Delray Beach, FL 33484.
Note: [***] Human Performance Laboratory, Florida Atlantic University, Boca Raton, FL.
Note: [****] Fitness Director, American Med-Care Center and Rehabilitation Facility, Palm Springs, FL.
Abstract: The purpose of this study was to assess the test-retest concentric reliability of the LIDO Active isokinetic dynamometer. Twenty subjects performed knee extension and flexion on two separate days, 48 hours apart. Peak torque (PT), average work (AW), average power (AP) and joint angle at peak torque (JA) were measured at 60, 180, 300 and 400 degrees per second (deg/sec). Pearson product moment (r) correlations between day 1 and day 2 for PT, AW, and AP ranged from 0.90 to 0.99. The JA r values ranged from 0.09 to 0.92. Extension paired t-tests revealed that mean values for PT, AW, AP and JA on day 1 were not significantly (p < 0.05) different from mean values on day 2, with the exception of AW at 300 deg/sec. Flexion t tests indicated that PT values were not different. Average work was significant at all velocities except 300 deg/sec, while AP and JA were significant at 180 and 60 deg/sec, respectively. It was concluded that the LIDO Active isokinetic dynamometer is a reliable instrument to assess concentric isokinetic strength.
Keywords: LIDO active, reliability, concentric mode
DOI: 10.3233/IES-1992-2407
Journal: Isokinetics and Exercise Science, vol. 2, no. 4, pp. 191-194, 1992
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]